FDA Medical Device Class II: GE Medical Systems, LLC – Senographe Pristina
Class II issued for Senographe Pristina. GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode w...
The FDA Center for Devices and Radiological Health (CDRH) has issued an official medical device recall determination for Senographe Pristina produced by GE Medical Systems, LLC.
Regulatory Recall Specification
- Recall Number: Z-0327-2027
- Recall Classification: Class II
- Firm Name: GE Medical Systems, LLC
- Current Status: Ongoing
Reason for Device Recall
GE HealthCare has become aware of a nonconformance to certain Senographe Pristina VIA systems. Specifically, in cases of 3D+2D manual mode where the acquisition configuration of the anode track/filter combination is different, it can result in the following: the collimator blades might be visible in the image, or the collimator might be open more than expected, and an incorrect collimator filter might be used with respect to the anode track selected This issue could result in a minor additional X-ray dose being delivered to the patient and obvious artifacts in the image (collimator blades).
Clinical & Corrective Actions
Healthcare facilities and clinical engineering departments are advised to review the affected serial numbers and follow the manufacturer's field safety notice instructions.